How to Read an HPLC Chromatogram: Peaks, RT and Area %

How to read a peptide HPLC chromatogram: retention time, void peak, peak area %, baseline integration, UV wavelength (214/220 vs 280 nm) and what it misses.

What an HPLC chromatogram shows

An HPLC chromatogram is a plot of detector response against time produced as a dissolved sample moves through a packed analytical column. The horizontal axis (x-axis) records time since injection, usually in minutes (Time [min]). The vertical axis (y-axis) records signal intensity — in ultraviolet (UV) absorbance detection, typically printed in milli-absorbance units (mAU) or millivolts (mV). A flat baseline means only mobile phase is passing through the detector flow cell; each peak marks a separated component or solvent disturbance reaching the detector.

Synthetic peptides are normally analysed by reversed-phase HPLC (RP-HPLC) on a silica stationary phase bonded with non-polar octadecyl (C18) or octyl (C8) chains, eluted with a water–acetonitrile gradient containing an ion-pairing modifier such as 0.1% trifluoroacetic acid (TFA). Because the stationary phase is hydrophobic, unretained salts and polar species leave the column earliest, the main peptide elutes at a characteristic retention time (tR) as acetonitrile rises, and more hydrophobic by-products elute later.

Anatomy of the trace and the peak table

Laboratory reports usually print the chromatogram trace directly above a software integration table (peak table). Reading the trace and the table together shows whether every visible feature on the curve was included in the calculation, rather than relying on the headline percentage alone.

Column labels vary slightly across chromatography data systems, but most peptide reports list RT [min], Width [min], Area, Height and Area%. Each part of the trace and table answers a distinct question.

Core features of a reversed-phase peptide HPLC chromatogram and peak table
Feature or columnWhere it appearsWhat it tells you
Void / injection disturbance (t0 or tM)Near the start of the run, often around 0.3–1.0 min depending on column volume and flow rate.Marks the column hold-up (void) time when unretained injection solvent and buffer ions pass the flow cell; normally excluded from peptide peak integration.
Main peak and retention time (tR)The dominant peak on the trace; its apex time is printed under RT [min].Elapsed time from injection to peak maximum. Reproducible under identical column, gradient, temperature and flow settings, though tR alone does not prove chemical identity.
Baseline and start/stop markersDrawn line beneath each integrated peak, bounded by tick marks or drop lines.Sets the zero-signal reference subtracted from the curve. Where the baseline begins, ends or splits adjacent peaks changes integrated peak areas.
Peak height vs peak areaSeparate numeric columns (Height in mAU; Area in mAU·s or software area units).Height reads signal at the apex; area integrates signal across the whole peak width and is standard for purity because band broadening lowers height more than area.
Area % (relative peak area)Final column of the peak table, summing to 100% across integrated peaks.A peak’s integrated area divided by the sum of all integrated peak areas, times 100. This relative figure is what peptide COAs report as HPLC purity.
Minor impurity peaks (early vs late)Smaller integrated peaks eluting before or after the main peptide peak.On a C18 gradient, early-eluting peaks are more hydrophilic (such as truncated or oxidized species); late-eluting peaks are more hydrophobic (such as side-chain adducts or aggregates).

How peak area % is calculated — and how integration affects it

Chromatographic purity on a peptide report is a relative area percentage, not a weight fraction: each peak’s Area is divided by the total Area (Sum) of all integrated peaks and multiplied by 100. Because the denominator includes only integrated peaks, two integration settings applied to the same raw data file can yield slightly different percentages without any change in the sample.

Baseline placement, blank exclusion and threshold settings drive most differences. Drawing a baseline too high under a small peak clips away part of its area; conversely, solvent-front spikes near t0 and gradient background peaks that also appear in a blank solvent injection should be excluded so artefacts are not counted as sample impurities. A high area-reject or slope threshold can likewise leave minor baseline-resolved peaks unintegrated.

Peak shape and axis scaling also deserve attention. Symmetrical Gaussian peaks are straightforward to integrate, whereas pronounced peak tailing (a drawn-out trailing edge, often from secondary silanol interactions) or peak fronting (a sloped leading edge, often from column mass overload) can hide co-eluting shoulders and shift baseline drop points. Check that the time axis starts at 0.0 min rather than being cropped tightly around the main peak, and treat a trace printed without its numeric peak table as incomplete.

Why detection wavelength matters (214/220 nm vs 280 nm)

An UV-Vis HPLC detector does not weigh molecules leaving the column; it measures light absorbance through a flow cell at a selected wavelength. Because every chromophore has its own molar absorptivity at a given wavelength, equal masses of two compounds rarely give equal peak areas.

Analytical HPLC of peptides routinely monitors 214 nm or 220 nm (within the 210–220 nm low-UV window above the ~190 nm cutoff of HPLC-grade acetonitrile), where the peptide bond itself absorbs strongly regardless of amino acid composition. By contrast, 280 nm — widely used to estimate total protein concentration — primarily measures the aromatic rings of tryptophan (Trp, molar extinction ~5500 M⁻¹ cm⁻¹) and tyrosine (Tyr, ~1490 M⁻¹ cm⁻¹), plus a minor contribution from disulfide cystine (~125 M⁻¹ cm⁻¹).

Changing the wavelength therefore changes relative peak area percentages whenever a peptide and an impurity carry different chromophores. A peptide sequence without Trp or Tyr gives a strong peak at 214–220 nm yet almost disappears at 254 nm or 280 nm, while a trace aromatic reagent by-product can look tiny at 214 nm and prominent at longer wavelengths. Always record the detection wavelength alongside any HPLC purity percentage.

Worked walkthrough: BPC 157 report BC10-20260908S6179

To see these elements on an actual file, open BPC 157 10 mg report BC10-20260908S6179 (sample BC10/BPC157, batch 5828704, received and analysed 2026-09-08, issued by TJQXGD) in our report library. Under HPLC-MS Details, the primary UV trace is labelled VWD1A, Wavelength=220 nm, plotted across 0.0–6.0 min on the horizontal axis and 0–800 mAU on the vertical axis, with a small unintegrated injection dip near 0.4 min.

Below the trace, the 220 nm peak table lists two manually integrated peaks (code MM m): the main peak at RT 3.132 min (Width 0.4333 min, Area 4766.9738, Height 754.2392, Area% 99.8779) and a minor late-eluting peak at RT 5.434 min (Width 0.2076 min, Area 5.8255, Height 1.1714, Area% 0.1221). Dividing 4766.9738 by Sum 4772.7993 gives 99.8779%, which appears in the summary table as Purity (LCMS/HPLC): 99.87% against specification ≥ 99.0%.

Page 2 of the PDF also illustrates why wavelength matters: a second channel, VWD1B, Wavelength=254 nm, shows the same peak at 3.135 min with a height of only 1.0684 mAU (Area 6.6508) — more than seven hundred times smaller than at 220 nm — because BPC-157 (G-E-P-P-P-G-K-P-A-D-D-A-G-L-V) contains no Trp, Tyr or Phe residues. At the same time, note what the printout omits: 5.000 µL injection volume and method file LCMS-02AB.amx are named, whereas stationary phase, column dimensions, mobile-phase gradient and a blank injection trace are not listed.

BPC 157 10 mg

Verify online: BC10-20260908S6179

BPC 157

Documentation

What an HPLC chromatogram cannot tell you

Even a full-window 214 nm or 220 nm trace with > 99% main-peak area describes only the share of UV-absorbing, column-resolved signal in that peak. Several critical properties are invisible to UV-HPLC alone and require orthogonal methods or batch records.

When reviewing a chromatogram on a peptide COA or third-party report, work through these six checks in order:

  • Check the axes and run window: confirm Time [min] begins at 0.0 min (showing the void region) rather than being cropped around the main peak, and note the mAU or mV scale.
  • Read the detection wavelength (such as 214 nm or 220 nm) so you know peptide bonds — not only aromatic side chains — were monitored.
  • Cross-check the trace against the peak table: verify that visible minor peaks are integrated, inspect baseline markers and integration codes, and confirm main Area divided by Sum equals Area%.
  • Record which method parameters are stated or absent (column chemistry, gradient, injection volume, blank subtraction) before comparing figures from different laboratories.
  • Combine HPLC purity with orthogonal tests on the same sample: mass spectrometry (MS) for molecular mass, plus peptide weight or net peptide content when weighing material for solutions.
  • Verify the report number on the issuer’s website and match the printed batch/lot number to your quotation and to the vial labels received.
What reversed-phase UV-HPLC misses and which test or record addresses each gap
Blind spotWhy UV-HPLC alone misses itWhat addresses it instead
Chemical identity, isomers and co-elutersTwo molecules that share the same retention time under one gradient merge into one peak; D/L enantiomers and some sequence inversions have matching hydrophobicity.Mass spectrometry (MS) for molecular mass, MS/MS or amino acid analysis for sequence, chiral chromatography for optical isomers, or an orthogonal HPLC gradient.
Salts, counter-ions and water (net peptide content)Water, inorganic salts, acetate and TFA do not absorb as peptide peaks at 214–220 nm or elute in the void volume, so a 99% pure powder often holds 70–90% peptide by weight.Net peptide content (elemental nitrogen or quantitative amino acid analysis), Karl Fischer titration for water, and ion chromatography for TFA or acetate.
Endotoxin and microbial contaminationBacterial lipopolysaccharides (endotoxins) and microbes occur at trace levels invisible on an analytical UV chromatogram.Limulus amebocyte lysate (LAL) or recombinant Factor C assay (reported in EU/vial or EU/mg) and microbiological testing.
Link between the tested vial and your shipmentA chromatogram characterizes only the sample solution injected on the analysis date.Online verification on the issuer’s website and lot-number matching across quote, report and received labels.

Chromatograms in our report library and lot documents

Every entry in our report library links to the complete source PDF or full-page scan alongside the issuer’s online verification record, allowing you to inspect the printed chromatogram, detection wavelength, integration table and mass spectrum directly.

The issuer (TJQXGD) states that the data come from a third-party partner laboratory, that neither it nor the laboratory holds CMA/CNAS accreditation, and that this document has no legal certification validity. The issuer (PRISTINBIO) states that the analysis is for information only and applies to the sample provided.

If your laboratory requires current-lot records, clarification of a peak table or method field, or testing through an accredited laboratory of your choice before ordering, include your catalog code, pack count and destination in a quote request.

Ask for the documents for the exact lot in your quote. A report describes only the batch it tested, so check that its lot number matches the lot you receive.

Request a quote

What is a lot number? Lot vs batch number explained

Chromatography and analytical references

For laboratory research. Not for human or animal use.

NORTHVERITY BIO / Materials

Catalog materials referenced in this guide

Inspect the compound profile, available pack specifications and published reference reports, or add an item to your quote list.

BPC-157BC5 · 5 mg × 10 vials · 1419.5 g/mol
View product
NORTHVERITY BIO / B2B procurement

Know what the document covers.

Ask for the documents for the exact lot in your quote. A report describes only the batch it tested, so check that its lot number matches the lot you receive.

Documentation guides

NORTHVERITY / B2B procurement

Let’s build your next inquiry.

One list. Your specifications. A direct conversation.

Request a quote
NORTHVERITY / INDEX

Search the collection

ProductsDocumentationGuides
NORTHVERITY / Image viewer

Image viewer

100%

NORTHVERITY / RFQ

Paste a product list

One SKU and pack quantity per line. Example: CU50, 2. Review before adding.

NORTHVERITY / PROCUREMENT

Your quote list

Prices are not listed on the site. Send your quote list and we confirm price, availability and delivery in a written quote. A quote list is not an order. Delivery costs and current-lot documents are confirmed separately.

Your quote list is stored on this browser for up to 30 days. Customer details are not saved in browser storage. Measurement choices are under Optional measurement at the foot of the page.

Continue to inquiry
NORTHVERITY / MATERIALS

Compare specifications

Prices are not listed on the site. Send your quote list and we confirm price, availability and delivery in a written quote. Each report covers only the sample and batch printed on it. Verify it on the issuer’s website and check that the lot number matches the goods you receive.

Optional measurement

Umami visit statistics (no cookies or visitor identifiers) are on unless you choose Essential only; first-party source measurement is optional. Separately, OpenAI ad measurement uses automatic matching: after you consent, its SDK may read supported contact fields and send hashed email, phone and name values, plus available location and browser data, for ad attribution. It reports page views and when you open WhatsApp, Telegram or email here or send an inquiry; message content is not sent. Google Ads conversion measurement reports the same actions, with Google advertising cookies that recognise visits from a Google ad; form contents and contact details are not sent to Google. In the US and Canada it is on unless you choose Essential only or Analytics only; in Quebec and elsewhere it sets no cookies unless you choose Analytics + automatic ad matching. This permission is optional; enquiries work without it. Privacy information

Saved selection

Only SKUs and pack quantities are stored. No contact fields or messages.

Your current selection will be replaced after confirmation.

WhatsAppChat for a quote